Front end x-ray beam position monitors at the Canadian Light Source
Date
2006-09-15
Authors
Journal Title
Journal ISSN
Volume Title
Publisher
ORCID
Type
Degree Level
Masters
Abstract
The development of X-ray Beam Position Monitors (XBPM) used on the Canadian Light Source front ends is described in this thesis, from the design concepts to the practical implementation and commissioning. Surveyed into position to provide a fiducialized point of origin for incoming synchrotron radiation, the primary purpose of the XBPM is to provide a measure of synchrotron beam motion. Currently XBPMs have been installed on three beamlines at the Canadian Light Source, a 2.9 GeV third generation synchrotron radiation source. Two of the XBPMs are comprised of chemical vapour deposition synthetic diamond blades coated with gold and installed on insertion device beamlines, while the third makes use of molybdenum blades for a dipole beamline. By incrementally scanning the blades of the XBPM through the synchrotron beam it is possible to determine the monitors' spatial resolution to beam motion. For the commissioned XBPM a typical spatial resolution of +/- 1 micron of beam motion was achieved; and the thermal power loading capacity has been tested to the 2/5 of maximum value. An independent white beam profiler, comprised of a converter crystal and image acquisition software, was constructed to corroborate the functionality of the XBPM.
Description
Keywords
x-ray beam position monitor, synchrotron radiation, front end
Citation
Degree
Master of Science (M.Sc.)
Department
Physics and Engineering Physics
Program